Cyclic Salt Corrosion Chamber Installation Physical Specification
Model Number | CCT-NC-20 | CCT-NC-30 | CCT-NC-40 | CCT-NC-90 | CCT-NC-110 |
Internal Capacity | 565 ltr. | 850 ltr. | 1130 ltr. | 2548 ltr. | 3115 ltr. |
20 cu.ft. | 30 cu.ft. | 40 cu.ft. | 90 cu.ft. | 110 cu.ft. | |
Test Space | .94 square meter | 1.40 square meter | 1.74 square meter | 2.21 square meter | 4.18 square meter |
Testing Plane | 10.1 square feet | 15.1 square feet | 18.7 square feet | 23.7 square feet | 45 square feet |
Internal Dimension | 1270 mm L (50”) | 1905 mm L (75”) | 1905 mm L (75”) | 1930 mm L (76”) | 2267 mm L (89.25”) |
L x W x H | 737 mm W (29”) | 737 mm W (29”) | 914 mm W (36”) | 1143 mm W (45”) | 1860 mm W (73.25”) |
635 mm H (25”) | 635 mm H (25”) | 635 mm H (25”) | 1155 mm H (45.5”) | 762 mm H (30”) | |
Total Chamber Volume | 982.5 ltr. | 1,472.5 ltr. | 1,925.5 ltr. | 3,822 ltr. | 4,870.5 ltr. |
34.7 cu.ft. | 52.0 cu.ft. | 68.1 cu.ft. | 135 cu.ft. | 172 cu.ft. | |
Floor Space | 2590 mm L (102”) | 3226 mm L (127”) | 3226 mm L (127”) | 3353 mm L (132”) | 3658 mm L (144”) |
Required | 2057 mm W (81”) | 2057 mm W (81”) | 2286 mm W (90”) | 2540 mm W (100”) | 3912 mm W (154”) |
External Dimensions | |||||
Left to Right | 2413 mm (95”) | 3048 mm (120”) | 3048 mm (120”) | 3353 mm (132”) | 3480 mm (137”) |
Front to Back with cover on and closed | 914 mm (36”) | 914 mm (36”) | 1092 mm (43”) | 1676 mm (66”) | 2286 mm (90”) |
Tall – Floor to top of cover (closed) | 1448 mm (57”) | 1448 mm (57”) | 1448 mm (57”) | 2235 mm (88”) | 1829 mm (72”) |
The inside dimensions, capacities, and space given here are “available testing space”. They do not include space below the false floor where the heaters are located or within the cover.
Temperature Ranges:
Ambient to 71 oC (160 oF)
Humidity:
95-100% During High Humidity from Ambient to 50 oC (122 oF) < 30% during Dry off cycles
Controllable Humidity available as an option upon request from 30-95% +/- 5% as limited by a 22oC dew point and lab conditions. NOT ALL TEMPERATURE AND HUMIDITY COMBINATIONS ARE ATTAINABLE! (Check with factory for details)
Solution Consumption during fogging tests only:
CCT-NC-20 approx. 10 gallons per day for D.I. water and salt solution if a fog only cycle is running and no cycling takes place.
CCT-NC-30, 40 & 90 approx. 15 gallons per day for D.I. water and salt solution if a fog only cycle is running and no cycling takes place.
NOTE: D.I. water consumption increases with cyclic testing when cycling from fogging to non-fogging tests.
Air Consumption:
Approximately 3.5 CFM, must be a constant 95 psi (6.55 Bar) supply of clean dry air or the machine will not run. D.I. Requirement: ASTM D1193 Type 4 water ONLY, must be a constant – 60 – 100 psi (4.5 – 6.9 Bar) supply of clean water or the machine will not run.
Sample Capacity:
CCT-NC-20 |
CCT-NC-30 |
CCT-NC-40 |
CCT-NC-90 |
|
4” x 12” panels |
160 |
240 |
360 |
390 |
3” x 6” panels |
190 |
280 |
380 |
416 |
Venting:
4.5” Diameter (114 mm)
Drain
Main drain is ¾” (19 mm) pipe drain that must be free flowing and open at all times.
There are multiple drain lines depending on options.
Model Number |
208/1/60 |
208/3/60 |
240/1/60 |
240/3/60 |
460/3/60 |
CCT-NC20 |
35 amps |
30 amps |
30 amps |
26 amps |
16 amps |
NC-20 with Immersion |
N/A |
68 amps |
N/A |
61 amps |
31 amps |
CCT-NC30 |
44 amps |
40 amps |
38 amps |
34 amps |
20 amps |
NC-30 with Immersion |
N/A |
77 amps |
N/A |
69 amps |
36 amps |
CCT-NC40 |
44 amps |
40 amps |
38 amps |
34 amps |
20 amps |
NC-40 with Immersion |
N/A |
77 amps |
N/A |
69 amps |
36 amps |
CCT-NC90 |
53 amps |
53 amps |
47 amps |
47 amps |
25 amps |
NC-90 with Immersion |
N/A |
165 amps |
N/A |
155 amps |
75 amps |
These amperage ratings are the actual amperage draw for the cabinet. 480 volt 3 phase available as an option at additional cost.