AUTO TECHNOLOGY CHAMBER INFORMATION OVERVIEW

Thin horizontal red-to-gray gradient bar
Auto Technology Legacy Steel Salt Fog chamber Legacy Steel Salt Fog For legacy QA and QC corrosion testing, especially where a lab already uses this traditional chamber style. Common for established salt fog, water fog and high-humidity methods. Common Industries: Automotive • Aerospace & Defense • Coatings & Metal Finishing • General Manufacturing Common Specifications: ASTM B117 • ISO 9227 • MIL-STD-810 Method 509.7 • ASTM D2247 • ASTM G85 A1
Control Types
Control Type Functions
Standard Controls Controls chamber temperature and bubble-tower heater temperature. Air pressure is adjusted separately. Fog baffles are manually adjustable.
Touchscreen / Computer Controls Controls chamber temperature, data logging and basic cyclic functions.
Customizations
Customization Control / Platform Limitation Applicable Standards / Test Types
Wet-Bottom / Humidity Pipe System Air bubbles through water in the chamber bottom. The lowered water jacket heats the air and creates condensation on the specimens. ASTM D2247 and applicable high-humidity / condensation procedures.
High-Temperature Package Raises chamber temperature for methods in the 50 °C range. ASTM B368 CASS, ISO 9227 CASS, ASTM G85 A2 / A3 where required.
SO2 Dosing Package Adds SO₂ dosing. Computer controls are preferred for dosing control. ASTM G85 A4 and applicable SO2 salt-fog procedures.
Horizontal Fog Dispersion Provides more usable test space for large or unusual specimens. Used when additional clear specimen space is required.
Applicable Specifications
Specification / Method Configuration
ASTM B117Base salt-fog configuration.
ISO 9227 NSSBase salt-fog configuration.
ISO 9227 AASSSalt-fog configuration suitable for acidified solution.
ASTM G85 Annex A1Acetic acid salt spray.
JIS Z 2371Salt-spray configuration.
GB/T 10125 NSS / AASSSalt-spray configuration.
ISO 7253Neutral salt-spray configuration.
IEC 60068-2-11Salt-mist configuration.
EIA-364-31Connector salt-spray configuration.
MIL-STD-810 Method 509.7Salt-fog configuration.
MIL-STD-202 Method 101Salt-atmosphere configuration.
MIL-STD-883 Method 1009Salt-atmosphere configuration.
MIL-STD-750 Method 1046.3Salt-atmosphere configuration.
ASTM B368 / ISO 9227 CASSHigh-Temperature Package for the required elevated chamber temperature.
ASTM G85 Annex A2Cyclic / temperature configuration.
ASTM G85 Annex A3Cyclic / temperature configuration.
ASTM G85 Annex A4SO2 dosing package.
ASTM D2247Wet-bottom / humidity-pipe configuration.
ASTM D1735Water fog using DI water in the solution tank and the chamber fog-generation system; formal humidity setup may also use the humidity pipe / wet-bottom arrangement as specified.
ISO 6270 humidity / condensation methodsHumidity configuration.
Add-Ons
Add-On Function
Bubble Tower DI Holding Tank / Hybrid Feed SystemSupplies the bubble tower when pressurized facility DI water is not available.
Shut Down TimerProvides timed shutoff of air to the bubble tower for unattended operation.
Sample Access PortsProvides openings for specimen wiring, tubing or other external connections.
Independent Data LoggerProvides additional recording of chamber-related temperatures and operating data.
Scrubber / Jet Exhaust AssemblyTreats chamber exhaust where direct exterior ventilation is unavailable.
Recirculating Scrubber / Jet Exhaust SystemRecirculates scrubber water instead of using continuous fresh-water flow.
Combination Jet Exhaust / Wet-Bottom Drain AssemblyCombines exhaust handling with the wet-bottom drain arrangement.
Power VentProvides powered exhaust assistance for longer vent runs.
External Condensate Collection PackageAllows collection-rate checks without opening the chamber cover.
Slotted Support Bars / Slotted Support RacksProvides organized support and spacing for test panels and specimens.
Heavy-Duty ShelvesSupports heavier or nonstandard specimens.
Custom Specimen FixturesAdapts the exposure zone to specific part geometry.
Corrosion / Mass-Loss CouponsProvides reference specimens for applicable chamber-verification or corrosion methods.
Calibration SupportProvides chamber-related calibration documentation and service.
NRTL Field Inspection LabelOptional NRTL field inspection and labeling, available at additional cost.
Auto Technology Advanced Accelerated Corrosion chamber Advanced Accelerated Corrosion For corrosion testing from simple salt fog to complex cyclic programs involving heat, immersion, spray, humidity control, climate transitions and freezing. Common Industries: Automotive OEM & Tier Suppliers • Transportation • Coatings & Metals • Aerospace • Industrial Common Specifications: SAE J2334 • GMW 14872 • GMW 3172 • Ford L-467 • VDA 233-102 • ISO 11997-3 • ASTM G85
Control Types
Control Type Example Functions
Touchscreen Controls Auto Technology Advanced Accelerated Corrosion chamber with touchscreen controls Controls temperature, salt fog, high humidity, timed cycles and direct solution spray / impingement.
PC Controls — Controls advanced RH and temperature transitions, recirculated spray, spray cooling, freezing, swaying spray, immersion and multiple solution sources.
Customizations
Customization Touchscreen PC Controls Applicable Standards / Test Types
Water Fog Uses DI water through the fog system. Uses DI water through the fog system. ASTM D1735 and other water-fog procedures.
Wet Bottom Heats water in the chamber bottom to create high humidity and condensation. Standard on most sizes; not standard on Size 110. Heats water in the chamber bottom to create high humidity and condensation. Standard on most sizes; not standard on Size 110. ASTM D2247 and related high-humidity / condensation exposures.
Humidity Control Controls defined RH setpoints for J2334-style testing. Controls defined RH setpoints for J2334-style testing. SAE J2334 and other cyclic methods requiring defined humidity setpoints without full climate-transition control.
Advanced Climate Control — Controls programmed RH and temperature transitions. VDA 233-102, ISO 11997-3 and other advanced cyclic methods requiring controlled temperature / RH transitions.
Direct Solution Spray / Impingement Direct solution application to specimens. Direct solution spray / impingement with PC sequencing. GMW 14872, SAE J2334 and related OEM cyclic procedures.
Recirculated Solution Spray — Collects and reuses sprayed solution. GMW 3172 and related procedures using recirculated solution spray.
Cooled Recirculated Spray — Cools solution before it is recirculated. GMW 3172 and related procedures requiring controlled or cooled recirculated spray solution.
Mechanical Cooling / Freezing — Adds below-ambient and freezing capability. VDA 233-102, ISO 11997-3 Cycle C and other procedures containing low-temperature / freezing stages.
LN2 Low-Temperature Assist — LN2-assisted low-temperature operation for applications requiring additional cooling capacity. Low-temperature cyclic corrosion procedures requiring the LN2-assisted configuration.
Swaying Spray — Moves the spray across the specimen area. Ford CETP 00.00-L-467 / L-467.
Immersion — Transfers solution from a separate tank into the chamber for immersion, then returns it to the tank. OEM / customer cyclic procedures combining immersion with other environmental stages.
Multiple Solution Sources — Allows spray from multiple solution sources. OEM / customer cyclic procedures requiring different solutions in separate spray stages.
Horizontal Dispersion Towers Horizontal fog-dispersion arrangement for additional usable specimen space; used on the Size 40 touchscreen configuration. Horizontal fog-dispersion arrangement for additional usable specimen space. ASTM B117, ASTM B368, ASTM D1735 and other fog methods using the horizontal dispersion arrangement.
SO2 Dosing — Adds sulfur-dioxide dosing under computer control. ASTM G85 Annex A4 and related SO2 corrosion cycles.
Applicable Specifications
Specification / Method Touchscreen PC Controls Configuration Notes
ASTM B117Yes.Yes.Salt fog is a base capability.
ISO 9227 NSS / AASSYes.Yes.Salt-fog configuration.
ASTM B368 / ISO 9227 CASSYes.Yes.Within the standard ambient-to-70 °C chamber temperature range.
ASTM G85 Annex A1Yes.Yes.Acetic acid salt spray.
ASTM G85 Annex A2Yes.Yes.Cyclic acidified salt spray.
ASTM G85 Annex A3Yes.Yes.SWAAT.
ASTM G85 Annex A5Yes.Yes.Fiberglass / Advanced Accelerated Corrosion only.
ASTM D2247Yes.Yes.Wet-bottom high-humidity / condensation exposure; wet bottom is not standard on Size 110.
ASTM D1735Yes.Yes.Water fog using DI water through the fog-generation system.
ISO 6270 humidity methodsMethod-dependent.Yes.Exact RH / condensation requirements determine hardware.
JIS Z 2371Yes.Yes.Salt-spray configuration.
IEC 60068-2-11Yes.Yes.Salt-mist configuration.
GMW 14872Yes.Yes.Direct solution spray / impingement with the required humidity and cycle functions.
SAE J2334Yes.Yes.Humidity Control option for defined RH thresholds; direct solution spray / impingement as required.
Ford CETP 00.00-L-467 / L-467—Yes.Swaying spray with the required programmed environmental cycle.
VDA 233-102—Yes.Advanced climate control; Cycle C uses mechanical cooling / freezing.
ISO 11997-3—Yes.Advanced climate control; Cycle C uses mechanical cooling / freezing.
ISO 11997-1Method dependent.Yes.Method-specific controls and hardware.
GMW 3172—Yes.Recirculated solution spray; cooled recirculated spray where required.
ISO 16701Method dependent.Yes.Cycle-specific controls and hardware.
ISO 14993Method dependent.Yes.Cycle-specific controls and hardware.
IEC 60068-2-52Method dependent.Yes.Cyclic salt mist with conditioning / dry stages.
Toyota cyclic corrosion proceduresMethod dependent.Yes.Exact procedure / revision determines spray, temperature and humidity hardware.
Volkswagen PV-series corrosion proceduresMethod dependent.Yes.Procedure-specific configuration.
Volvo cyclic corrosion proceduresMethod dependent.Yes.Procedure-specific configuration.
Hyundai / Kia cyclic corrosion proceduresMethod dependent.Yes.Procedure-specific configuration.
Customer-specific OEM cyclic proceduresMethod dependent.Yes.Built to the required sequence and hardware.
Add-Ons
Add-On Availability / Function
Support Angles / Support Rods / Slotted Support Bars / Support Racks / Heavy-Duty ShelvesSpecimen support, positioning and loading.
Custom Specimen Racks / FixturesAdapts the chamber to specimen geometry.
Wet Bulb Humidity MeasurementMeasures wet-bulb temperature and allows calculated RH monitoring from wet-bulb and dry-bulb temperature. It is a measurement option, not active RH control.
Access Ports / FeedthroughsProvides access for wiring, tubing, sensors or instrumentation.
Automatic Cover LiftersAssists with cover opening and closing.
Mixing Tank Convenience PackageSimplifies solution mixing, storage and transfer.
Additional Bulk Solution ReservoirsAdds solution storage capacity or separate convenience reservoirs.
Automatic Solution FillReduces manual solution replenishment.
Water Scrubber / Recirculating Scrubber / Jet ExhaustHandles chamber exhaust and salt carryover.
Power Exhaust VentProvides powered assistance for facility exhaust runs.
Data LoggingAvailable according to the installed control platform.
Remote Monitoring / Network AccessAvailable on PC-controlled configurations and other supported control packages.
Collection Funnels & CylindersSupports fog collection-rate checks.
Corrosion / Mass-Loss CouponsReference specimens for applicable methods.
External Condensate CollectionCollects chamber condensate externally.
Calibration PackagesProvides applicable chamber calibration documentation and services.
Maintenance SupportFactory maintenance and service support.
NRTL Field Inspection LabelOptional NRTL field inspection and labeling, available at additional cost.
Auto Technology mixed flowing gas chamber Mixed Flowing Gas (MFG) For modern gas-corrosion testing, especially in electronics, connectors, contacts and other components sensitive to low-level atmospheric pollutants. Common Industries: Electronics • Electrical Connectors & Contacts • Telecommunications • Automotive Electronics • Industrial Reliability Common Specifications: IEC 60068-2-60 • EIA-364-65 • ASTM B827 • ASTM B845 • Telcordia GR-1217 • Battelle Class I–IV
Control Types
Control Type Functions
Integrated PC Controls Controls recipes, data logging, gas delivery, airflow, air exchange, temperature and humidity.
Customizations
Customization Control / Configuration Applicable Standards / Test Types
Gas Type / Mix Configured around the pollutant gases required by the test. Supported gases include H2S, SO2, NO2 and Cl2. IEC 60068-2-60, EIA-364-65, ASTM B827, ASTM B845, Battelle and customer-specific MFG procedures.
Gas Analyzers Measures chamber gas concentration for closed-loop control. Up to four analyzer inputs. MFG procedures requiring measured working-volume gas concentration and closed-loop concentration control.
Mass-Flow Controls Controls pollutant-gas flow. MFG procedures requiring controlled low-level gas delivery.
Humidity Control Controls chamber RH. IEC 60068-2-60, EIA-364-65, ASTM B827, ASTM B845, Battelle and customer-specific MFG procedures.
Temperature Control Controls exposure temperature using the dual-chamber design. IEC 60068-2-60, EIA-364-65, ASTM B827, ASTM B845, Battelle and customer-specific MFG procedures.
Permeation Tube System Generates pollutant gas with permeation tubes. MFG procedures using permeation-tube gas generation.
Applicable Specifications
Specification / Method Configuration Notes
IEC 60068-2-60Mixed flowing gas exposure with the required gas mixture, concentration, temperature, RH and air exchange.
EIA-364-65Mixed flowing gas testing of connectors / sockets; environmental class and control-coupon requirements determine setup.
ASTM B827Practice for conducting mixed flowing gas corrosion exposures.
ASTM B845Mixed flowing gas test conditions for electrical contacts / devices.
Telcordia GR-1217Applicable mixed-gas / corrosive-atmosphere configurations as required by the program.
Battelle Class IConfigured to the selected Battelle environment.
Battelle Class IIConfigured to the selected Battelle environment.
Battelle Class IIIConfigured to the selected Battelle environment.
Battelle Class IVConfigured to the selected Battelle environment.
Customer-specific electronics pollutant-gas environmentsConfigured to the required gas species, concentrations, airflow, air exchange, temperature and RH.
Add-Ons
Add-On Function
Custom Shelves / Electronics Racks / FixturesSupports specific specimen layouts.
Wiring Ports / Additional FeedthroughsProvides access for powered or instrumented specimens.
Gas Detection / Facility Safety InterfacesSupports facility gas-monitoring requirements.
Remote Monitoring / Network AccessProvides remote access to supported control and logging functions.
Coupon / Corrosivity Monitoring HardwareSupports atmosphere verification with applicable control coupons or monitoring methods.
Mixing FansProvides additional chamber-air circulation where specified.
Custom Cabinet DimensionsAdapts usable test volume to the specimen load.
NRTL Field Inspection LabelOptional NRTL field inspection and labeling, available at additional cost.
Auto Technology Gas Dosing chamber Gas Dosing For moist-gas corrosion testing using controlled gas dosing. Common for SO₂, H₂S and similar gas exposure methods. Common Industries: Electronics & Electrical • Metals & Coatings • Connectors & Contacts • Industrial Materials Common Specifications: DIN 50018 • ASTM G87 • ISO 3231 • IEC 60068-2-42 • IEC 60068-2-43
Control Types
Control Type Functions
Command Controls Physical controls for temperature and gas dosing.
Customizations
Customization Control / Configuration Applicable Standards / Test Types
Number of Gases Configured for one to four gases. DIN 50018, ASTM G87, ISO 3231, IEC 60068-2-42, IEC 60068-2-43 and customer-defined gas-dosing procedures.
Applicable Specifications
Specification / Method Configuration Notes
DIN 50018SO2 condensation / Kesternich-type moist-gas exposure.
ASTM G87Moist SO2 exposure.
ISO 3231Humid atmospheres containing SO2.
IEC 60068-2-42SO2 exposure where the required gas delivery is compatible with the G4 dosing approach.
IEC 60068-2-43H2S exposure where the required gas delivery is compatible with the G4 dosing approach.
Customer-defined moist-gas proceduresConfigured around the required gas species, humidity, temperature, flow and exposure time.
Customer-defined gas-dosing exposuresConfigured around one to four gas inputs and the required dosing conditions.
Add-Ons
Add-On Function
Custom Specimen Racks / SupportsSupports specific specimen layouts.
Access FittingsProvides additional tubing or instrumentation access.
Gas Supply InterfacesAdapts the chamber to the facility gas-supply arrangement.
Facility Exhaust ConnectionsConnects the chamber to the required exhaust system.
External Gas DetectionSupports facility gas-monitoring requirements.
Custom Chamber DimensionsAdapts usable test volume to the specimen load.
NRTL Field Inspection LabelOptional NRTL field inspection and labeling, available at additional cost.
Auto Technology simple immersion chamber Simple Immersion For continuous or cyclic immersion in water, saltwater or other compatible solutions. Used for coatings, metals, plastics and finished components. Common Industries: Coatings • Metals • Pipeline & Infrastructure • Energy • Plastics & Materials Common Specifications: ASTM D870 • ASTM G31 • ISO 2812-2 • ISO 175 • Customer-Specific Immersion Methods
Control Types
Control Type Functions
Dedicated Immersion Controls Controls solution temperature, heaters, liquid level and other process functions required by the configured tank.
Programmable Cyclic Controls Used on cyclic fill-and-drain systems to control pumps, valves, immersion duration, drain duration and cycle count.
Customizations
Customization Control / Design Limitation Applicable Standards / Test Types
Continuous vs. Cyclic Immersion Configured for continuous immersion or repeated fill-and-drain cycles. Continuous immersion methods or customer-specific cyclic immersion procedures.
Solution Heating / Temperature Control Dedicated heater and RTD feedback configured for the required solution temperature. ASTM D870, ASTM G31, ISO 2812-2, ISO 175 and customer procedures specifying solution temperature.
Air Sparging / Agitation Dedicated airflow / agitation hardware where required. Customer and material-test methods requiring liquid movement or aeration.
Cyclic Fill & Drain Requires programmable controls, reservoir, pumps, valves and drain sequencing. Customer-specific cyclic immersion and repeated immersion / drain procedures.
Multiple Solutions / Reservoirs Custom plumbing and control arrangement based on chemistry and sequence. Customer procedures requiring different liquid exposures.
Test-Specific Immersion Depth / Tank Geometry Tank dimensions and liquid level are built around the specimen and required exposure depth. Application-specific.
Solution-Compatible Construction Tank, pump, seal, heater and plumbing materials must be selected for the test chemistry. Application-specific based on liquid chemistry.
Applicable Specifications
Specification / Method Configuration Notes
ASTM D870Water immersion of coatings.
ASTM G31Laboratory immersion corrosion testing of metals.
ISO 2812-2Water immersion.
ISO 175Effects of liquid chemicals on plastics.
Customer-specific water immersionTank size, solution temperature and specimen loading are configured to the procedure.
Customer-specific saltwater immersionMaterials of construction and solution handling are configured to the chemistry.
Customer-specific cyclic immersionRequires the appropriate fill / drain reservoir, pumps, valves and programmable controls.
Internal durability / repeated immersion-drain proceduresConfigured to the required dwell, drain and cycle sequence.
Add-Ons
Add-On Function
Custom Specimen Racks / BasketsSupports loading and removal of test specimens.
Heavy-Part Supports / Loading FixturesSupports larger or heavier components.
Custom CoversProvides application-specific tank access and closure.
Access PortsProvides access for sensors, tubing or instrumentation.
Convenience Drains / Overflow PlumbingImproves fill, drain and overflow handling.
External Transfer ConnectionsSupports external liquid transfer or facility plumbing.
NRTL Field Inspection LabelOptional NRTL field inspection and labeling, available at additional cost.
Auto Technology Century Series spray chamber Legacy Spray For rain, spray and water-ingress testing. Used for automotive parts, lighting, enclosures, electronics and other products exposed to direct water. Common Industries: Automotive • Lighting • Electronics & Electrical • Enclosures • Industrial Equipment Common Specifications: IEC 60068-2-18 Method R / Rb • JIS D 0203 • Customer & OEM Water-Ingress Procedures
Control Types
Control Type Functions
Basic ON / OFF Controls Used for simple fixed spray exposures requiring basic pump / spray operation and timing.
PLC / Touchscreen Controls Used for timed nozzle events, oscillating spray, specimen motion, heated water and multi-step sequencing.
Customizations
Customization Control / Configuration Applicable Standards / Test Types
Nozzle Type / Quantity / Location Fixed or sequenced nozzle arrangements. Method-specific rain, spray, wash and water-ingress procedures.
Droplet Characteristics / Spray Pattern / Angle Set by nozzle type, geometry and water delivery. Applicable IEC, JIS, SAE and customer/OEM water-exposure procedures.
Pressure / Flow Configuration Pump, regulator, nozzles and instrumentation are selected for the required water pressure and flow. Water-ingress, rain, wash and direct-spray procedures.
Impact Velocity / Nozzle Distance Set by nozzle type, pressure, flow and distance. Procedures defining water impact conditions or nozzle-to-specimen geometry.
Water Temperature Control Requires heater / temperature-control hardware and programmable control when temperature changes are part of a sequence. Procedures specifying controlled spray-water temperature.
Oscillating / Moving Spray Requires powered motion and PLC / touchscreen sequencing. Methods requiring moving spray coverage.
Multiple Spray Zones / Intermittent or Sequential Spray Uses PLC / touchscreen controls for multiple zones or timed sequences. Customer/OEM procedures with staged or zoned water application.
Multiple Test Liquids Uses separate reservoirs, pumps and plumbing for multiple liquids. Customer-defined procedures using more than one test liquid.
Specimen Rotation / Powered Motion Requires programmable control when rotation, reversing or powered specimen operation is part of the exposure. JIS D 0203 and other water-ingress / spray methods requiring controlled specimen motion.
Applicable Specifications
Specification / Method Configuration Notes
IEC 60068-2-18 Test RbImpacting-water exposure. The specific Rb method and severity determine the spray apparatus.
JIS D 0203Applicable R1 / R2 / S1 / S2 rain / spray configurations as required by the selected method.
Customer-specific rain testingNozzle, pressure, flow, water temperature, angle, duration and specimen motion configured to the procedure.
OEM water-intrusion proceduresBuilt to the required water delivery and specimen motion.
Direct water-spray durability proceduresBuilt to the required spray apparatus and sequence.
Add-Ons
Add-On Function
Adjustable Specimen Racks / Custom FixturesSupports specimen positioning and loading.
Access Ports / FeedthroughsProvides access for wiring, sensors or instrumentation.
Larger Convenience ReservoirsAdds water storage capacity.
Additional Drain ConnectionsImproves water removal and facility plumbing.
Observation FeaturesProvides additional visibility of the specimen during exposure where configured.
Convenience PlumbingImproves filling, draining and connection to facility water systems.
NRTL Field Inspection LabelOptional NRTL field inspection and labeling, available at additional cost.