C SERIES CORROSION TEST & SALT SPRAY CHAMBER

Thin horizontal red-to-gray gradient bar

The Auto Technology C Series salt spray chamber runs NSS, AASS, and CASS salt spray tests, humidity exposures, and ASTM G85 cyclic corrosion tests for coated panels, plated parts, and industrial components. Choose Size 20 or Size 40, then select the monitoring, exhaust, and specimen supports your test program requires.

Auto Technology C Series salt spray chamber with viewing lid and touchscreen controls
  • NSS / AASS / CASS
    ASTM B117 & ISO 9227
  • ASTM G85 A1 / A2 / A3 / A5
  • Salt fog, humidity & cyclic operation
  • Size 20 & Size 40

Tests the C Series Can Run

The C Series supports the following salt spray, cyclic corrosion, and humidity methods. Chamber setup and test solution depend on the standard and revision.

NSS — Neutral Salt Spray

ASTM B117 / ISO 9227 NSS. Continuous neutral salt spray for metals, coatings, and plated finishes.

ASTM B117 laboratory testing

CASS — Copper-Accelerated Acetic Acid Salt Spray

ISO 9227 CASS. Copper-accelerated acetic-acid salt spray testing.

Discuss a dedicated CASS chamber with ATC. Residual copper contamination can affect later NSS testing.

ISO 9227 CASS test requirements

Common Applications

Choose Your Chamber Size

Choose by part dimensions, quantity, and the racks or supports you need. Allow room around the chamber for utilities, exhaust, and service access.

Size 20

20 ft³ Test volume: 594 L

C Series Size 20 chamber drawing
Panel capacity
Approximately 160 small, flat panels using 4 racks
Chamber footprint
95 × 43 in
Fog dispersion towers
One
View Size 20 Drawings

Size 40

40 ft³ Test volume: 1,189 L

C Series Size 40 chamber drawing
Panel capacity
Approximately 360 small, flat panels using 9 rack positions
Chamber footprint
120 × 43 in
Fog dispersion towers
Two
View Size 40 Drawings

Size 20 and Size 40 are nominal size designations. Panel capacities assume optional slotted trays and small, flat specimens. These estimates use the 40-slot test panel tray. Larger panels and close spacing can cause shadowing and reduce the practical load. Custom fixtures are available for larger components and assemblies. Confirm the final footprint against the quoted configuration and drawings.

Usable Test Space

ATC bases chamber capacity on the specimen exposure area below the lid. The sloped lid provides clearance and directs condensate away from samples; its volume is excluded from the stated test space.

Send your largest part dimensions and the number of specimens per test. We can check the loading arrangement and recommend racks or shelves.

C Series Size 20 chamber loaded with corrosion test panels
Size 20 shown with optional panel supports.
Compare Usable Chamber Volume

Compare the volume below the lid when evaluating specimen capacity. ATC lists 594 L for Size 20 and 1,189 L for Size 40. The leading competitor lists approximately 500 L and 900 L excluding the lid for its 600 and 1100 models.

ChamberTest Volume Excluding LidSolution Reservoir
ATC C Series Size 20594 L60 US gal
Q-FOG 600Approximately 500 LApproximately 32 US gal
ATC C Series Size 401,189 L60 US gal
Q-FOG 1100Approximately 900 LApproximately 32 US gal

The leading competitor also lists approximately 600 L and 1,100 L when lid volume is included. Source: Leading competitor Specification Bulletin LF-8110.21. Volumes are approximate. Part dimensions, spacing, and racks determine how many specimens fit.

ATC Size 20 and Size 40 comparison of specimen volume and solution reservoir capacity
ATC chamber-size and reservoir comparison. Click to enlarge.

What Comes Standard

Fiberglass Cabinet

Molded, corrosion-resistant cabinet with a viewing lid.

Touchscreen Controller

6-inch color display for stored methods, programming, manual functions, and chamber status.

60-Gallon Reservoir

Built-in 60-gallon solution tank for fewer refills during extended tests.

Fog Generation

Venturi atomizer, bubble tower, and baffled fog dispersion system.

Heating & Humidity

Chamber heating and wet-bottom humidity capability. Exposure-zone temperature: ambient to 70°C; high-humidity operation: approximately 95–100% RH.

Lower Shelving

Standard lower-level shelving for specimen support.

Alarms & Shutdowns

Chamber alarms and protective shutdown functions.

C Series Size 20 standard interior with lower shelving
Standard Size 20 interior. Size 40 uses two dispersion towers.

Racks & Specimen Supports

Lower shelving is included. Select optional support rods, slotted bars, panel racks, or mid-level shelves to suit your specimens and the required exposure angle.

Review sample-handling options

Field Upgrades

The C Series can be field-upgraded for additional cyclic tests. Available upgrades depend on the chamber’s existing controls and hardware. Include any future test requirements when requesting a quote.

Options & Accessories

Include the options you need in your quote request.

Measurement & Documentation

Data Logging

Saves chamber and bubble-tower temperatures to the controller memory card. Also records wet-bulb temperature when the Wet Bulb Option is fitted.

Part 701350

Wet Bulb Option

Adds wet-bulb temperature measurement to check chamber humidity.

Part 710675

Calibration Certificate

NIST-traceable calibration certificate for your quality records.

Part 701900

Solution Preparation

Mixing Tank Convenience Package

C Series mixing tank convenience package

Prepare test solution at the chamber with an air sparger for mixing, an overflow port, and manual valves for controlling air and water.

Part 700825

Exhaust

Select the exhaust arrangement for your facility before ordering.

Standard Scrubber

Non-recirculating scrubber using a continuous supply of pressurized water discharged to drain. Discuss this option when direct exterior venting is impractical.

Part C700200

Recirculating Scrubber

Recirculating Scrubber

Jet exhaust system that reuses scrubber water, reducing water consumption compared with a flow-through scrubber.

Part C260315

Power Exhaust Vent Kit

Power Exhaust Vent Kit

Required for exhaust runs over 15 ft. ATC can also review runs with multiple elbows or other restrictions.

Part 701000

Sample Handling

Choose racks and supports to suit the chamber size, specimen weight, and required exposure angle.

Standard Size 20 Interior

C Series Size 20 standard chamber interior with lower shelving

Standard chamber interior with included lower shelving, before adding optional racks and supports.

Support Angles

Support Angles

Support racks, collectors, and other accessories.

Support Rods

Support Rods

Suspend parts from ½-in-diameter rods.

Slotted Bars

Slotted Bars

Hold panels at 15° or 6° from vertical.

Panel Racks

Panel Racks

8.5 × 23 in trays with 40 slots.

Mid-Level Shelves

Mid-Level Shelves

Additional support for parts and slotted bars.

Racks, Bars & Shelves — Part Numbers
Part NumberAccessorySize / UseOrder
700400Support Angles – Size 20Size 20; supports panel racks and collection accessories.View product
700405Support Angles – Size 40Size 40; supports panel racks and collection accessories.View product
700420Support Rods – Size 20Size 20; ½-in-diameter rods for hanging parts.View product
700425Support Rods – Size 40Size 40; ½-in-diameter rods for hanging parts.View product
70041015° Slotted Support Bar – Size 20Size 20; 3 × 6 in and 4 × 12 in panels at 15° from vertical.View product
70041515° Slotted Support Bar – Size 40Size 40; 3 × 6 in and 4 × 12 in panels at 15° from vertical.View product
C262880-S6° Slotted Support Bar – 29 in29-in polypropylene bar; 6° from vertical.View product
C262890-S6° Slotted Support Bar – 41 in41-in polypropylene bar; 6° from vertical.View product
C100324Slotted Support Rack8.5 × 23 in tray; two rows of 20 slots.View product
700430Heavy Duty Mid-Level Shelf – Size 20Composite mid-level shelf for Size 20.Browse store
700432Heavy Duty Mid-Level Shelf – Size 40Composite mid-level shelf for Size 40.View product

Place slotted polypropylene bars on heavy-duty shelves for added support and to reduce sagging.

Test Supplies & Startup Items

Order premixed solutions, salt, DI water, and setup supplies as needed for your test program.

Solutions, salt, DI water, and startup supplies
Part NumberItemDescriptionOrder
C173300ASTM B117 / ISO 9227 Salt Solution5% sodium chloride solution, 55-gallon drum.View product
C173305ASTM Type IV DI WaterASTM D1193 Type IV DI water, 55-gallon drum.View product
C173273-CCertified ASTM B117 SaltCertified 99.9% sodium chloride for solution preparation.View product
C173330ASTM G85 Annex A1 SolutionPremixed solution for ASTM G85 A1.View product
C173332ASTM G85 Annex A2 SolutionPremixed solution for ASTM G85 A2.View product
C173331ASTM G85 Annex A3 SolutionPremixed acidified synthetic seawater solution for ASTM G85 A3.View product
C173334ASTM D5894 / G85 Annex A5 SolutionDilute electrolyte solution for ASTM G85 A5.View product
C264000pH Meter KitpH measurement kit for solution checks.View product
701500Start-Up KitStartup supplies for CCT / NC chambers.View product
703100Installation KitInstallation components for CCT / NC chambers.View product

Training, Maintenance & Proficiency Testing

Operator Training

Remote instruction on setup, operation, maintenance, and troubleshooting.

Part C910001

Maintenance Kit

Replacement components for routine preventive maintenance.

Part C920000

Browse replacement parts and testing supplies

Installation Assistance & Service

ATC can help plan chamber placement and utilities, provide installation and startup assistance, and support the chamber with calibration, repairs, preventive maintenance, training, and replacement parts.

Chamber installation, calibration & repair services

Need a test solution that is not listed above? Ask about CASS solution and other method-specific supplies when requesting your quote.

How the C Series Works

During conventional salt fog operation, compressed air and prepared solution meet at the atomizer. The fog then passes through a baffled tower before reaching the specimens.

  1. Compressed AirClean, dry air enters the chamber.
  2. Bubble TowerAir is heated and humidified.
  3. Venturi AtomizerConditioned air draws and atomizes solution.
  4. Dispersion TowerA baffle removes large droplets.
  5. Exposure ZoneFog surrounds the specimens at the set temperature.

This flow describes conventional salt fog. ASTM G85 A5 uses dry-air fogging with the bubble tower bypassed; setup depends on the selected test method.

Measuring Fog Collection

Operators use collection funnels and graduated cylinders to measure fog fallout and check the collected solution. The controller manages temperature and the programmed test sequence.

Humidity Operation

For water fog, the chamber atomizes DI water. For condensing humidity, the wet bottom and submerged heater generate water vapor that condenses on specimen surfaces.

Water fog demonstration, shown at 1.5× speed.

Fog Delivery & Specimen Clearance

The dispersion tower carries fog above the loading area, where the baffle helps remove large droplets before fog reaches the specimens. This arrangement keeps the atomization point away from test surfaces and provides clearance around the loading area. Check the drawing and proposed rack layout for large parts, tall panels, or assemblies.

ATC illustration comparing baffled fog delivery and specimen clearance
Baffled fog delivery and specimen-clearance illustration. Click to enlarge.

Solution Temperature Before Atomization

The reservoir sits within the heated chamber environment, so the stored solution approaches chamber temperature before it reaches the atomizer.

Both C Series sizes have a 60-gallon reservoir, compared with the approximately 32-gallon reservoir listed for Q-FOG 600 and 1100 models. That is about 88% more stored solution; actual refill intervals depend on fog output, the test cycle, and operating conditions.

ATC illustration of internal reservoir placement and solution temperature
Solution-reservoir placement and temperature illustration. Click to enlarge.

C Series Touchscreen Controls

C Series touchscreen menu with Select Test to Run highlighted

Operators use the touchscreen to select a method, start a test, and check chamber status.

  • Preloaded test methods
  • Custom programs and stored sequences
  • Manual operation for setup and checks
  • Live chamber status
  • Alarm display

Need test records? Add the optional data-logging package for temperature records that can be transferred to a computer.

For bubble-tower pressure or RH-probe logging, see the options for an X Series cyclic corrosion chamber. C Series temperature logging is described in the data-logging FAQ.

Installation & Facility Requirements

Confirm utility connections and exhaust routing before delivery. Final requirements depend on chamber size and configuration.

Electrical

208, 240, or 480 V configurations. Confirm phase, load, and disconnect requirements against your quote and drawings.

Compressed Air

65–85 psi, clean and dry. Supply through ⅜-in O.D. poly tubing.

DI Water

Pressurized DI water, typically 15–30 psi, through ⅜-in O.D. poly tubing. Specify water quality for the test method, including ASTM D1193 Type IV where required.

Drain & Exhaust

1-in gravity drain and 2-in PVC exhaust connection. Power exhaust is required for runs over 15 ft.

Electrical loads and disconnect requirements
SizePart NumberSupply ConfigurationMaximum Current
20790020208 or 240 V; 1Ø / 3Ø configurations25.9 A
20790021480 V / 3Ø12.5 A
40790040208 or 240 V; 1Ø / 3Ø configurations34.2 A
40790041480 V / 3Ø16.7 A

Size the electrical circuit using the voltage, phase, and current specified on the factory drawings for your chamber.

  • A qualified electrician must provide a lockable fused disconnect with slow-blow fuses, in full operator view.
  • Mount the disconnect separately from the chamber cabinet.
  • Size conduit and cable for the specified voltage and load.
  • Provide noise/surge suppression where supply quality requires it.
Drain, exhaust, and facility details
  • Use 1-in I.D. tubing on the chamber drain hose barb. Discharge to an open drain; do not hard-plumb the drain connection.
  • Route exhaust directly outdoors wherever possible. Use a dedicated vent rather than a shared instrument duct or plenum.
  • Support vent piping, maintain proper pitch, and leave rear-service access.
  • Provide upstream filtration and drying for contaminated plant air. Purge air uses the compressed-air supply.
  • Place the chamber on a flat, level floor with adequate load capacity and a nearby floor drain.

Room Requirements

Room temperature: 20–25°C (68–77°F). Room humidity: 45 ±10% RH, non-condensing. Keep the chamber environment above freezing and below 60% RH.

Corrosion chamber exhaust vent and gravity drain connections

Frequently Asked Questions

Are salt spray and salt fog chambers different?

Salt spray chamber, salt fog chamber, and salt spray cabinet are common names for equipment that exposes specimens to an atomized salt solution. In ASTM B117 testing, the intended exposure is a fog atmosphere, with direct spray kept off the specimens. See the ASTM B117 apparatus requirements for the method’s chamber requirements.

Can the C Series run ASTM G85 Annex A5?

Yes, with the A5 fog/dry configuration. A5 uses dry-air fogging with the bubble tower bypassed.

Can I switch between salt spray and humidity testing?

Yes. The C Series supports water fog and wet-bottom condensing humidity. Clean the chamber before changing from salt solution to DI water, then set it up for the next method. Add the Wet Bulb Option when you need to measure humidity.

How do I choose between Size 20 and Size 40?

Size 40 provides more specimen space and uses two fog dispersion towers. Choose based on your largest part, quantity per test, support arrangement, and available floor space. The size cards include panel-capacity estimates and drawing links.

Does it need an exterior exhaust connection?

Direct exterior venting is preferred. If that is impractical, ask ATC to review a standard or recirculating scrubber arrangement. Exhaust runs over 15 ft require the Power Exhaust Vent Kit. Confirm routing before ordering.

What data can the controller log?

The optional data-logging package records chamber temperature and bubble-tower temperature to the touchscreen controller’s memory card. It also records wet-bulb temperature when the Wet Bulb Option is installed. Transfer the files to a computer for test records and reporting.

How is the C Series different from the M Series?

The C Series uses a molded fiberglass cabinet, internal heating, and standard touchscreen controls. The M Series uses a steel cabinet with water-jacket heating and Command Controls, with a touchscreen available as an option. Compare the required test cycle and configuration in the C & M Series selection guide.

When should I choose an X Series chamber?

Choose the X Series for cycles requiring recirculated spray, cooled spray, humidity ramps, or broader temperature and humidity control. ATC can review OEM cycles, immersion requirements, and low-temperature stages against your specification.

Why does the chamber use a baffled fog tower?

The tower separates atomization from the specimen area. Its baffle reduces large droplets and direct impingement before fog enters the exposure zone. Specimens must still be positioned to allow fog circulation and prevent runoff from one sample reaching another.

Why is the solution reservoir inside the chamber?

Keeping solution within the heated environment helps bring it toward chamber temperature before atomization. See the reservoir illustration and fog-generation explanation for the layout and the 60-gallon capacity comparison.

Can salt spray hours be converted into years of outdoor service?

There is no universal conversion. Salt spray testing is used for comparisons and specification requirements. A service-life prediction needs evidence for the material, coating, and actual exposure environment.

What should I send for a chamber quote?

Send the standard and revision, largest specimen dimensions, quantity per test, available floor space, electrical supply, and any monitoring or reporting requirements. Include your exhaust route or a room sketch if you need installation planning assistance.

Request a C Series Quote

Send us your test standard, specimen dimensions, and quantity. We will help select the chamber size, supports, and options for your lab.

Request a C Series Quote