IEC 60068-2-52 CYCLIC CORROSION TEST CHAMBER – ATC X2-52

Thin horizontal red-to-gray gradient bar

The Auto Technology X2-52 cyclic salt mist chamber runs IEC 60068-2-52 exposures in one cabinet. Baffled fog towers deliver salt mist, and Advanced Climate Control sets the temperature and relative humidity for the stages in your test method. Specify the IEC method when ordering. Methods 7 and 8 also require timed transitions between mist, dry and humid conditions.

Size 20 chamber shown with a single baffled fog tower and optional upgraded RTDs.
Size 20 chamber shown with a single baffled fog tower and optional upgraded RTDs.

Baffled salt fog towers

The atomizer uses compressed air conditioned in the heated bubble tower to disperse salt solution. A baffle removes larger droplets before the mist settles on the specimens. ATC sizes the tower arrangement for the exposure area.

During the salt mist stage, the chamber operates at 35 ±2 °C. The collection target is 1–2 mL/h per 80 cm² of collection area. A pressure transducer monitors the bubble-tower air pressure.

Fog tower with a horizontal baffle, typically used in Size 40 and larger chambers.
Fog tower with a horizontal baffle, typically used in Size 40 and larger chambers.

Advanced Climate Control

A circulation fan moves air around the specimens. A temperature RTD and a shielded RH probe measure conditions in the exposure zone. The computer control uses those readings to adjust heating, cool-mist humidification and drying.

Advanced Climate Control requires computer-controlled pneumatic cover lifters, which keep the lid closed while the blower runs.

Advanced Climate Control humidification and air circulation beside the specimens
Size 110 exposure zone showing the cool-mist nozzle, RH sensor and circulation fan.

Computer control and data logging

Set the stage order, temperature, relative humidity, hold times and repeats for the specified IEC method. The computer control operates the chamber equipment and records measured temperature and RH so the lab can review the transitions as well as the holds.

Laptop displaying the chamber control software beneath the raised console cover
Chamber computer control with the console cover raised.

Which IEC 60068-2-52 method does your lab run?

Use the method named in the product specification. All eight methods apply salt mist at 35 ±2 °C. For Methods 1–6, the humid hold is 40 ±2 °C and 93 ±3% RH.

MethodProgrammed exposureTest duration or cycle length
12 h salt mist, then 6 days 22 h humid.4 weeks
22 h salt mist, then 22 h humid.3 days
3–6Four daily mist/humidity cycles as in Method 2, followed by 3 days at 23 ±2 °C and 50 ±5% RH.Method 3: 1 week · Method 4: 2 weeks · Method 5: 4 weeks · Method 6: 8 weeks
72 h neutral salt mist, 4 h dry at 60 ±2 °C and below 30% RH, then 2 h humid at 50 ±2 °C and above 95% RH.8 h per cycle
8Same timing and climate profile as Method 7, using acidified salt solution.8 h per cycle

Transition times for Methods 7 and 8

Specify the required mist-to-dry transition (within 30 minutes or 30–60 minutes) and dry-to-humid transition (within 15 minutes or 15–30 minutes). The return from humid to salt mist must take no more than 30 minutes. Each transition counts toward the stage being entered. Include the mounted load weight so ATC can confirm transition performance for your setup.

IEC recommends a dedicated chamber for Method 8 because residual acidified solution can affect subsequent neutral salt mist tests.

View the Size 20 IEC 60068-2-52 chamber

Chamber sizes and drawings

Send the dimensions of your largest loaded fixture and the number of specimens per run so ATC can confirm the chamber size and collector placement.

SizeNominal capacity (ft³)Approx. volume (L)Internal L × W × H (in.)Base L × W (in.)
202059450 × 29 × 2595 × 43
303089175 × 29 × 25120 × 43
40391,10675 × 36 × 25120 × 50
90902,55076 × 45 × 45132 × 68
1101123,19489 × 73 × 30137 × 85
View Size 20 drawings
Base length: 95 in.
View Size 40 drawings
Base length: 120 in.
View Size 110 drawings
Base length: 137 in.

X Series chamber sizes and configurations

Electrical service and installation

The table shows standard X Series current ratings at 60 Hz. Use the factory electrical drawing to confirm the amperage and feeds for the quoted chamber.

Size208 V / 1 ph208 V / 3 ph240 V / 1 ph240 V / 3 ph460 V / 3 ph
2035 A30 A30 A26 A16 A
3044 A40 A38 A34 A20 A
4044 A40 A38 A34 A20 A
9053 A53 A47 A47 A25 A
110See drawingSee drawingSee drawingSee drawingSee drawing

For room conditions and utility connections, see the chamber installation guide.

X Series accessories and test solutions

View chamber accessories and test solutions

X2-52 frequently asked questions

Can one X2-52 run all eight IEC test methods?

The X2-52 can be configured for any of the eight methods. Specify the methods you plan to run and any transition times in your test plan. IEC recommends a dedicated chamber for Method 8 to avoid acidified solution residue affecting later neutral tests.

What should I provide for a Methods 7 or 8 quote?

Send the transition times in your test plan and the total weight of the parts and fixtures in a full load so ATC can review the configuration.

Request an X2-52 quote

Include the IEC test method and revision, required transition times where applicable, largest mounted part dimensions, quantity per run, combined weight of specimens and fixtures, and available voltage, phase and frequency.