SAE J2334 CYCLIC CORROSION TEST CHAMBER – ATC X2334
Specify the salt-application method when ordering an Auto Technology X2334 SAE J2334 cyclic corrosion chamber: direct spray, air-atomized fog, or immersion. All three configurations run the same humid and dry stages. Your choice determines the equipment used to keep the parts wet during the 15-minute salt stage.

Which wetting method does your lab use?
Start with the customer’s test plan. For comparison with earlier results or another lab, confirm which salt-application method was used. Spray, fog and immersion can produce different results even when the temperature, humidity and cycle times match.
Direct spray

Adjustable nozzles deliver solution directly to the parts. Position the heads around the loaded fixture so every required surface stays visibly wet for the full 15 minutes. Spray may be periodic or continuous. Use enough flow to maintain wetness without disturbing coatings or corrosion products.
Air-atomized fog

Additional fog towers distribute atomized solution through the exposure zone. This configuration supports the higher J2334 collection rate of 2–4 mL/h. Allow space for the collectors with the parts installed and document the measured rate.
Immersion

A pump transfers solution from an external tank into the chamber. The parts stay immersed for the full 15 minutes, then the solution drains back before drying. Specify the required immersion depth with the parts mounted. Immersion was the basis for most of the standard’s development work.
Need more than one method? ATC can install all three wetting systems in one X2334. List each method your lab needs in the RFQ so the quote includes the corresponding hardware.
Immersion equipment and fill sequence
SAE J2334 daily cycle
| Stage | Duration | Exposure |
|---|---|---|
| Humid | 6 hours | 50 ±2 °C and 100% RH. Specimens remain visibly moist or wet. |
| Salt application | 15 minutes | At ambient conditions, the selected system keeps the specimens visibly wet throughout the interval. |
| Dry | 17 hours 45 minutes | 60 ±2 °C and 50 ±5% RH, with air circulation around the parts. |
The ramp into drying counts toward the dry stage. The ramp back to humidity counts toward the next humid stage. For five-day operation, maintain dry-stage conditions over the weekend. For seven-day operation, continue the daily cycle.
Each wetting method uses the same mixed salt solution: 0.5% sodium chloride, 0.1% calcium chloride and 0.075% sodium bicarbonate in deionized water.
Advanced Climate Control and computer control
All three X2334 configurations include Advanced Climate Control for the humid stage and the 50% RH hold during drying. The package requires computer-controlled pneumatic cover lifters to hold the lid closed while the blowers run.
The computer control sequences the humid exposure, operates the installed wetting system, and records chamber temperature and relative humidity throughout the cycle.

Chamber sizes and drawings
Size for the largest loaded fixture. Direct spray needs a clear path to the exposed surfaces, fog needs space for collectors, and immersion needs enough operating depth to cover the parts. The base dimensions cover the chamber cabinet; the immersion configuration also requires floor space for its external tank.
| Size | Nominal capacity (ft³) | Approx. volume (L) | Internal L × W × H (in.) | Base L × W (in.) |
|---|---|---|---|---|
| 20 | 20 | 594 | 50 × 29 × 25 | 95 × 43 |
| 30 | 30 | 891 | 75 × 29 × 25 | 120 × 43 |
| 40 | 39 | 1,106 | 75 × 36 × 25 | 120 × 50 |
| 90 | 90 | 2,550 | 76 × 45 × 45 | 132 × 68 |
| 110 | 112 | 3,194 | 89 × 73 × 30 | 137 × 85 |
Installation and utilities
Provide a noncondensing laboratory environment at 25 ±5 °C and 45 ±10% RH for Advanced Climate Control.
| Utility | Connection or service |
|---|---|
| DI water | ASTM D1193 Type IV, 20–40 psi. |
| Compressed air | Clean, dry, oil-free air at 65–85 psi at the chamber inlet during operation. |
| Exhaust | Vent outdoors or connect to a suitable scrubber. |
| Drain | Open floor drain for the 1-inch gravity drain hose. |
| Power | Dedicated circuit and fused disconnect sized for the ordered configuration. |
Electrical service by wetting method and chamber size
Current ratings in amperes at 60 Hz. A dash indicates that no rating is listed for that supply.
| Configuration | Size | 208 V / 1 ph | 208 V / 3 ph | 240 V / 1 ph | 240 V / 3 ph | 460 V / 3 ph |
|---|---|---|---|---|---|---|
| Direct spray or fog | 20 | 35 | 30 | 30 | 26 | 16 |
| Direct spray or fog | 30 or 40 | 44 | 40 | 38 | 34 | 20 |
| Direct spray or fog | 90 | 53 | 53 | 47 | 47 | 25 |
| Immersion | 20 | — | 68 | — | 61 | 31 |
| Immersion | 30 or 40 | — | 77 | — | 69 | 36 |
| Immersion | 90 | — | 165 | — | 155 | 75 |
For Size 110 and combined wetting packages, obtain the connected load from the quoted factory drawing. Use the final drawing to size the disconnect for the ordered chamber.
Utility Connections & Exhaust Details
- DI water and compressed air: ⅜-in OD tubing on Sizes 20, 30, 40 and 90; ½-in OD on Size 110. Install an industrial oil/water separator upstream of the air connection.
- Exhaust: 2-in CPVC on Sizes 20–90 and 4-in CPVC on Size 110. Support and slope piping for condensate drainage. Exterior runs longer than 15 ft require additional exhaust assistance; confirm the arrangement with ATC.
- Drain: Route the 1-in gravity drain hose to an open floor drain.
- Laboratory: Maintain 25 °C ±5 °C for Advanced Climate Control and noncondensing room conditions.
Shut down the chamber if a required utility is interrupted. See the installation guide and use the factory drawing for the final layout.
X Series accessories and test solutions
View chamber accessories and test solutions
X2334 questions
Which salt application system should I order?
Specify direct spray, air-atomized fog or immersion to match the wetting method in your test program. For immersion, provide the dimensions of the loaded fixture so ATC can size the tank and confirm the fill sequence.
Does the wetting method change the climate controls?
All three configurations include Advanced Climate Control and computer-controlled pneumatic cover lifters for the humidity and drying stages.
Request an X2334 quote
Identify the wetting method your customer requires: direct spray, air-atomized fog or immersion. If the callout only says “SAE J2334,” send the drawing or test plan with your inquiry. Include the largest part’s mounted dimensions, quantity per run, fixture weight and available electrical service. For immersion, include the depth needed to cover the mounted parts.