VDA 233-102 CYCLIC CORROSION TEST CHAMBER – ATC X233-102
The Auto Technology X233-102 cyclic corrosion chamber runs all seven days of the VDA 233-102 exposure in one chamber. Every week includes all three required daily cycles: A, B and C. Additional fog towers deliver the Cycle A salt exposure, Advanced Climate Control runs the temperature and humidity profiles, and included mechanical refrigeration takes the chamber to −15 °C for Cycle C.

Mechanical cooling for the −15 °C stage
Cycle C includes a five-hour hold at −15 °C. The refrigeration system cools the exposure zone on the programmed ramp, holds the cold stage, then the chamber returns to the warmer profile. The parts stay on their fixtures throughout.
ATC sizes the refrigeration system for the cabinet and the combined mass of the specimens and fixtures. Coated evaporator coils protect the heat-exchange surfaces exposed to the corrosive chamber environment.

Additional fog towers for Cycle A
Cycle A calls for three hours of atomized fog using 1% sodium chloride solution. The extra towers distribute fog across the usable test space. The VDA collection target is 3.0 ±1.0 mL/h per collector with approximately 80 cm² of collecting area.
Distribution matters as much as output: each collector must meet the specified range. The VDA check is performed in the empty chamber over at least three hours.

Advanced Climate Control and computer control
The computer control sequences the three daily profiles and records the measured chamber temperature and relative humidity. Advanced Climate Control regulates humidity through the specified warm ramps and holds. RH control is suspended during the subzero portion of Cycle C.
Advanced Climate Control requires computer-controlled pneumatic cover lifters, which hold the lid closed while the blower runs. Chamber venting equalizes pressure during fog generation, heating and cooling.

Required seven-day VDA 233-102 sequence
The computer control runs one 24-hour cycle each day in the order shown. All seven days make up one complete VDA exposure week.
| Day | Required 24-hour cycle | Exposure |
|---|---|---|
| 1 | Cycle B · Climate and evaluation | Salt-free temperature and humidity cycling with an evaluation period. |
| 2 | Cycle A · Salt fog and climate | Three hours of 1% NaCl fog within the programmed temperature and humidity cycle. |
| 3 | Cycle C · Climate and freezing | Programmed cooldown, a five-hour hold at −15 °C and a return to warmer conditions, without salt application. |
| 4 | Cycle A · Salt fog and climate | Three hours of 1% NaCl fog within the programmed temperature and humidity cycle. |
| 5 | Cycle B · Climate and evaluation | Salt-free temperature and humidity cycling with an evaluation period. |
| 6 | Cycle B · Climate and evaluation | Salt-free temperature and humidity cycling with an evaluation period. |
| 7 | Cycle A · Salt fog and climate | Three hours of 1% NaCl fog within the programmed temperature and humidity cycle. |
ISO 11997-3 uses the same basic climate sequence but specifies a different fog-distribution check. Identify the standard called out on the drawing or test plan when requesting a quote.
Chamber sizes and drawings
Choose from five chamber sizes based on the loaded fixture dimensions and specimen quantity. Base footprints cover the cabinet only. Include the cooling unit in the installed layout.
| Size | Nominal capacity (ft³) | Approx. volume (L) | Internal L × W × H (in.) | Cabinet only, L × W (in.) |
|---|---|---|---|---|
| 20 | 20 | 594 | 50 × 29 × 25 | 95 × 43 |
| 30 | 30 | 891 | 75 × 29 × 25 | 120 × 43 |
| 40 | 39 | 1,106 | 75 × 36 × 25 | 120 × 50 |
| 90 | 90 | 2,550 | 76 × 45 × 45 | 132 × 68 |
| 110 | 112 | 3,194 | 89 × 73 × 30 | 137 × 85 |
X Series chamber sizes and configurations · Size 40 VDA chamber specifications
Size 40 layouts with mechanical cooling
The Mini Cool example measures 157.50 × 57.25 in. across the chamber and cooling assembly. Its movable solution tank is shown separately. The 10 HP example places the cooling unit behind the chamber and spans 122.19 × 119.40 in. as drawn. Allow additional service and utility access for either layout. For other sizes, use the installation drawing for the selected cooling package.
Installation and utilities
Mechanical cooling is included with the X233-102. Plan floor space and electrical service from the factory drawings for the chamber and cooling package together. Confirm the condenser arrangement before routing utilities: a water-cooled unit needs a separate process-water supply and return; a remote air-cooled unit needs an outdoor condenser location and a route for the refrigerant lines.
Provide a noncondensing laboratory environment at 25 ±5 °C and 45 ±10% RH for Advanced Climate Control.
| Utility | Connection or service |
|---|---|
| DI water | ASTM D1193 Type IV, 20–40 psi. |
| Compressed air | Clean, dry, oil-free air at 65–85 psi at the chamber inlet during operation. |
| Exhaust | Vent outdoors or connect to a suitable scrubber. |
| Condenser process water | A water-cooled refrigeration unit needs a facility process-water supply and return separate from the chamber DI feed. The Size 40 Mini Cool example has ¾-in NPT inlet and outlet ports and specifies 7 US gpm at 75 °F inlet water or 12 US gpm at 85 °F. Confirm flow, temperature and connections for the selected unit. A remote outdoor air-cooled condenser rejects heat outdoors and requires no condenser water. |
| Drain | Open floor drain for the 1-inch gravity drain hose. |
| Power | Confirm incoming voltage, phase, frequency, full-load current and disconnect requirements for the complete ordered system, including the refrigeration package. |
Size 20, 40 and 110 electrical examples
The switch rating, equipment fuse or breaker rating, and full-load current describe different electrical characteristics. These examples apply to the individual builds shown. Use the approved electrical drawing and nameplate for the ordered chamber and refrigeration package to size facility service.
| Size | Incoming supply (3-phase) | Customer disconnect | Equipment protection / load shown | Drawing |
|---|---|---|---|---|
| 20 | 480 V, 60 Hz | 60 A | 50 A fuses | 49039-01, 66036-01 |
| 20 | 380 V, 60 Hz | 60 A | 50 A breaker; 39.7 A full-load current shown | 79235-01 |
| 40 | 240 V, 60 Hz | 200 A | 100 A fuses | 49347-01 |
| 40 | 380 V, 50 Hz | 100 A | 70 A fuses on the 480 V transformer secondary | 57658-01 |
| 40 | 480 V, 60 Hz | 100 A | 70 A fuses | 58162-01, 60010-01 |
| 110 | 480 V, 60 Hz | 100 A | 90 A fuses; 71.6 A full-load current shown | 58704-01, 61294-01 |
Utility Connections & Exhaust Details
- DI water and compressed air: ⅜-in OD tubing on Sizes 20, 30, 40 and 90; ½-in OD on Size 110. Install an industrial oil/water separator upstream of the air connection.
- Exhaust: 2-in CPVC on Sizes 20–90 and 4-in CPVC on Size 110. Support and slope piping for condensate drainage. Exterior runs longer than 15 ft require additional exhaust assistance; confirm the arrangement with ATC.
- Drain: Route the 1-in gravity drain hose to an open floor drain.
- Laboratory: Maintain 25 °C ±5 °C for Advanced Climate Control and noncondensing room conditions.
Shut down the chamber if a required utility is interrupted, including loss of condenser-water flow on a water-cooled build. See the installation guide and use the factory drawing for the final layout.
X Series accessories and test solutions
View chamber accessories and test solutions
X233-102 questions
Do I select Cycle A, B or C when ordering?
The chamber runs all three as one seven-day sequence. Cycle A includes salt fog, and Cycle C includes the −15 °C freeze.
What does facilities need before installation?
Use the configuration drawing for the cabinet and refrigeration unit together. Confirm the electrical feeds and amperage for the quoted equipment. A water-cooled condenser also requires a process-water supply and return.
Request an X233-102 quote
Send the applicable standard, largest mounted part dimensions, quantity per run and total weight of the parts and fixtures. Include the available electrical service and whether your facility has process water for a water-cooled condenser or needs remote air cooling.