FORD L-467 CYCLIC CORROSION TEST CHAMBER – X467
The Auto Technology X467 cyclic corrosion chamber runs the complete Ford L-467 weekly exposure for CETP 00.00-L-467 and TM 00.00-L-467. A motorized spray bar applies salt solution across the parts. Advanced Climate Control brings the chamber back to wet conditions, warms the parts under high humidity and controls the two-hour drying ramp.

Motorized moving spray bar
The overhead bar sweeps overlapping fan sprays across the loaded fixture. It delivers fresh 0.5% sodium chloride solution directly to the parts at intervals during the six-hour wet stage. Ford targets an average downfall of 5 L/m² over the full six hours. Spent solution drains away after application.

Advanced Climate Control for temperature and humidity
At the start of the wet stage, the standard auxiliary blower draws in cooler laboratory air and exhausts warm chamber air. Cool mist helps cool the parts and maintain wetness. The program brings the chamber from 50 °C to below 30 °C within 30 minutes and to 25 °C within one hour.
Specify optional mechanical refrigeration only if the lab cannot keep its room temperature low enough for the standard blower to meet the Ford cooldown times.
After six hours, the chamber warms from 25 to 40 °C in 30 minutes at about 95% RH, keeping the parts wet. Over the next two hours, temperature rises from 40 to 50 °C while RH falls from 95% to 70%, allowing moisture in joints and crevices to dry gradually.
Advanced Climate Control requires computer-controlled pneumatic cover lifters, which keep the lid closed while the blower runs.

Preprogrammed Ford cycles
Select the preprogrammed Ford L-467 cycle using the chamber’s computer control. It sequences the spray bar and climate stages and records temperature and RH setpoints alongside measured conditions in the exposure zone.

Watch the X Series chamber walkthrough
Ford L-467 seven-day test cycle
The weekly program runs five daily wetting and controlled-climate cycles, followed by two days at 50 °C and 70% RH.
| Period | Stage | Duration | Nominal conditions |
|---|---|---|---|
| Days 1–5 | Wet stage | 6 hours | 25 °C; intermittent direct applications of 0.5% NaCl solution to maintain wet specimens. |
| Days 1–5 | Wet-to-dry transition | 2.5 hours | 25 to 40 °C at about 95% RH in 30 minutes, then a two-hour linear ramp to 50 °C and 70% RH. |
| Days 1–5 | Controlled climate | 15.5 hours | 50 °C and 70% RH. |
| Days 6–7 | Weekend hold | 48 hours | Continuous 50 °C and 70% RH; no scheduled weekday salt applications. |
Modified interior-body exposure
Ford also defines a modified interior-body exposure in which the normal climate sequence continues but salt is applied only on Mondays. Use the applicable Ford procedure and test plan to select the salt-application schedule and test duration.
Chamber sizes and drawings
Select a size that clears the moving spray bar with the tallest part mounted on its fixture. Check that the spray can reach each required surface with the full load in place.
| Size | Nominal capacity (ft³) | Approx. volume (L) | Internal L × W × H (in.) | Base L × W (in.) |
|---|---|---|---|---|
| 20 | 20 | 594 | 50 × 29 × 25 | 95 × 43 |
| 30 | 30 | 891 | 75 × 29 × 25 | 120 × 43 |
| 40 | 39 | 1,106 | 75 × 36 × 25 | 120 × 50 |
| 90 | 90 | 2,550 | 76 × 45 × 45 | 132 × 68 |
| 110 | 112 | 3,194 | 89 × 73 × 30 | 137 × 85 |
Electrical service and installation
Standard X Series current ratings at 60 Hz are shown below. An X467 equipped with mechanical refrigeration draws more current. Use the factory electrical drawing for the amperage and feeds on the quoted system.
| Size | 208 V / 1 ph | 208 V / 3 ph | 240 V / 1 ph | 240 V / 3 ph | 460 V / 3 ph |
|---|---|---|---|---|---|
| 20 | 35 A | 30 A | 30 A | 26 A | 16 A |
| 30 | 44 A | 40 A | 38 A | 34 A | 20 A |
| 40 | 44 A | 40 A | 38 A | 34 A | 20 A |
| 90 | 53 A | 53 A | 47 A | 47 A | 25 A |
| 110 | See drawing | See drawing | See drawing | See drawing | See drawing |
For room conditions and utility connections, see the chamber installation guide.
X Series accessories and test solutions
View chamber accessories and test solutions
Ford X467 frequently asked questions
Does the X467 need mechanical refrigeration?
The standard auxiliary blower uses laboratory air for cooldown. Specify mechanical refrigeration only if the lab cannot maintain the room temperature needed to meet the Ford cooldown times.
Can the X467 run ASTM B117 salt fog?
Yes. The X467 includes conventional salt fog for ASTM B117. The Ford L-467 cycle uses the moving spray bar to apply salt solution directly to the parts.
Request an X467 quote
Send the Ford test callout, part quantity, mounted dimensions, loaded fixture height and total load weight. Include the operating temperature and RH in your lab and the available electrical service. A fixture drawing or photo helps us check spray-bar clearance and coverage.