FORD L-467 CYCLIC CORROSION TEST CHAMBER – X467

Thin horizontal red-to-gray gradient bar

The Auto Technology X467 cyclic corrosion chamber runs the complete Ford L-467 weekly exposure for CETP 00.00-L-467 and TM 00.00-L-467. A motorized spray bar applies salt solution across the parts. Advanced Climate Control brings the chamber back to wet conditions, warms the parts under high humidity and controls the two-hour drying ramp.

Open X467 chamber with overhead Ford spray bar and optional perimeter spray heads
X467 with the overhead moving spray bar. The additional perimeter spray heads shown are optional.

Motorized moving spray bar

The overhead bar sweeps overlapping fan sprays across the loaded fixture. It delivers fresh 0.5% sodium chloride solution directly to the parts at intervals during the six-hour wet stage. Ford targets an average downfall of 5 L/m² over the full six hours. Spent solution drains away after application.

The Ford moving spray bar spans the exposure zone, with four nozzles spaced along its length.
The Ford moving spray bar spans the exposure zone, with four nozzles spaced along its length.

Advanced Climate Control for temperature and humidity

At the start of the wet stage, the standard auxiliary blower draws in cooler laboratory air and exhausts warm chamber air. Cool mist helps cool the parts and maintain wetness. The program brings the chamber from 50 °C to below 30 °C within 30 minutes and to 25 °C within one hour.

Specify optional mechanical refrigeration only if the lab cannot keep its room temperature low enough for the standard blower to meet the Ford cooldown times.

After six hours, the chamber warms from 25 to 40 °C in 30 minutes at about 95% RH, keeping the parts wet. Over the next two hours, temperature rises from 40 to 50 °C while RH falls from 95% to 70%, allowing moisture in joints and crevices to dry gradually.

Advanced Climate Control requires computer-controlled pneumatic cover lifters, which keep the lid closed while the blower runs.

Advanced Climate Control humidification and air circulation beside the specimens
Size 110 exposure zone showing the cool-mist nozzle, RH sensor and circulation fan.
Pneumatic cover lifter on an X Series corrosion chamber
Pneumatic cylinder for the computer-controlled cover lifter.
Auxiliary blower and vent on the X Series chamber
Standard auxiliary blower and vent for room-air cooling.

Preprogrammed Ford cycles

Select the preprogrammed Ford L-467 cycle using the chamber’s computer control. It sequences the spray bar and climate stages and records temperature and RH setpoints alongside measured conditions in the exposure zone.

X Series computer displaying the chamber software
X Series computer control interface.
Watch the X Series chamber walkthrough

Ford L-467 seven-day test cycle

The weekly program runs five daily wetting and controlled-climate cycles, followed by two days at 50 °C and 70% RH.

PeriodStageDurationNominal conditions
Days 1–5Wet stage6 hours25 °C; intermittent direct applications of 0.5% NaCl solution to maintain wet specimens.
Days 1–5Wet-to-dry transition2.5 hours25 to 40 °C at about 95% RH in 30 minutes, then a two-hour linear ramp to 50 °C and 70% RH.
Days 1–5Controlled climate15.5 hours50 °C and 70% RH.
Days 6–7Weekend hold48 hoursContinuous 50 °C and 70% RH; no scheduled weekday salt applications.
Modified interior-body exposure

Ford also defines a modified interior-body exposure in which the normal climate sequence continues but salt is applied only on Mondays. Use the applicable Ford procedure and test plan to select the salt-application schedule and test duration.

Ford L-467 test procedure and laboratory testing services

Chamber sizes and drawings

Select a size that clears the moving spray bar with the tallest part mounted on its fixture. Check that the spray can reach each required surface with the full load in place.

SizeNominal capacity (ft³)Approx. volume (L)Internal L × W × H (in.)Base L × W (in.)
202059450 × 29 × 2595 × 43
303089175 × 29 × 25120 × 43
40391,10675 × 36 × 25120 × 50
90902,55076 × 45 × 45132 × 68
1101123,19489 × 73 × 30137 × 85
View Size 20 drawings
Base length: 95 in.
View Size 40 drawings
Base length: 120 in.
View Size 110 drawings
Base length: 137 in.

X Series chamber sizes and configurations

Electrical service and installation

Standard X Series current ratings at 60 Hz are shown below. An X467 equipped with mechanical refrigeration draws more current. Use the factory electrical drawing for the amperage and feeds on the quoted system.

Size208 V / 1 ph208 V / 3 ph240 V / 1 ph240 V / 3 ph460 V / 3 ph
2035 A30 A30 A26 A16 A
3044 A40 A38 A34 A20 A
4044 A40 A38 A34 A20 A
9053 A53 A47 A47 A25 A
110See drawingSee drawingSee drawingSee drawingSee drawing

For room conditions and utility connections, see the chamber installation guide.

X Series accessories and test solutions

View chamber accessories and test solutions

Ford X467 frequently asked questions

Does the X467 need mechanical refrigeration?

The standard auxiliary blower uses laboratory air for cooldown. Specify mechanical refrigeration only if the lab cannot maintain the room temperature needed to meet the Ford cooldown times.

Can the X467 run ASTM B117 salt fog?

Yes. The X467 includes conventional salt fog for ASTM B117. The Ford L-467 cycle uses the moving spray bar to apply salt solution directly to the parts.

Request an X467 quote

Send the Ford test callout, part quantity, mounted dimensions, loaded fixture height and total load weight. Include the operating temperature and RH in your lab and the available electrical service. A fixture drawing or photo helps us check spray-bar clearance and coverage.