MIL-STD-810 Method 509.7 Salt Fog Chamber – C Series Size 20
The Auto Technology C Series Size 20 is a touchscreen-controlled salt fog chamber for MIL-STD-810 Method 509.7. The chamber produces the required neutral salt fog at 35 ±2 °C and can automatically transition to an ambient-air drying phase.
- Internal Volume: Approximately 20 ft³ (594 L)
- Exposure: 24-hour fog/24-hour dry cycling or 48-hour fog/48-hour dry exposure
- Fog Requirement: 1-3 mL/80 cm²/hour using 5 ±1% NaCl solution
- Controls: 6-inch color touchscreen with stored test sequencing
- Solution Capacity: Integrated 60-gallon reservoir
The Method 509.7 drying phase requires standard ambient temperature below 50% RH. Laboratory conditions must support that requirement.
Description
MIL-STD-810 Method 509.7 Salt Fog Chamber
The Auto Technology C Series Size 20 is configured for the neutral salt fog and ambient drying conditions used by MIL-STD-810H Method 509.7. The method is used to evaluate protective coatings and finishes and to identify physical, electrical, and corrosion problems associated with salt deposits.
The C Series combines atomized salt fog, chamber heating, touchscreen sequencing, and automatic ambient-air transitions in one corrosion-resistant cabinet.
Method 509.7 is a screening procedure: MIL-STD-810H does not present salt fog exposure as a direct predictor of coating service life. Test duration, item configuration, operational checks, and any variations must be established in the approved test plan.
C Series Size 20
| Usable exposure volume | Approximately 20 ft³ / 594 L |
|---|---|
| Base dimensions | 95 × 43 in / 242 × 110 cm |
| Controls | 6-inch color touchscreen |
| Solution reservoir | Integrated 60-gallon reservoir |
| Fog delivery | Venturi atomizer with baffled dispersion tower |

Method 509.7 Exposure Sequence
| Step | Condition |
|---|---|
| Preconditioning | Hold the test item at 35 °C for at least two hours before introducing salt fog. |
| Fog period | Continuously atomize the specified salt solution for 24 hours, or for the duration established by the test plan. |
| Dry period | Expose the item to standard ambient temperature below 50% RH for 24 hours. |
| Repeat | Repeat the fog and dry periods at least once, producing two wet periods and two dry periods. |
MIL-STD-810H also permits a 48-hour salt fog exposure followed by 48 hours of drying when the alternating 24-hour sequence is not used.
Salt Fog Conditions
| Chamber temperature | 35 ±2 °C |
|---|---|
| Salt concentration | 5 ±1% NaCl by weight |
| Collected-solution pH | 6.5-7.2 at 35 ±2 °C |
| Fallout rate | 1-3 mL/80 cm²/hour, based on at least 16 hours |
| Collectors | At least two, positioned nearest to and farthest from the nozzle |
| Air movement | Minimal during fog exposure |
Fog Generation and Chamber Interior

- Venturi atomizer produces a finely divided salt fog.
- Baffled dispersion reduces large droplets and direct specimen impingement.
- Humidifying tower conditions the compressed air used for atomization.
- Solution-contact components are constructed from corrosion-resistant materials.
- Exposed solution drains from the chamber and is not returned to the reservoir.
Touchscreen Cycle Control

The touchscreen sequences chamber heating, salt fog, and ambient-air drying phases. Operators can select stored programs, monitor chamber status, view alarms, and access manual functions.
Optional data logging captures chamber-performance information on the controller memory card. Independent calibrated instruments may also be used to satisfy laboratory reporting requirements.
Drying-condition requirement: The C Series introduces laboratory air during the dry phase. The laboratory must maintain standard ambient temperature and less than 50% RH for the chamber to reproduce the specified Method 509.7 drying condition.
Specify the MIL-STD-810 revision, exposure sequence, test-item dimensions, orientation, loading quantity, operational checks, and required data records when requesting a Method 509.7 Size 20 configuration.